Transparent
Calipers drawn on the image, heights in millimetres, the percentage lost and the rule that set the grade. Nothing rests on a score no one can inspect.
AI fracture detection on the CT scans you already take
Spine Sight AI reads the spine on every chest and abdominal CT, measures each vertebra, grades compression fractures, and puts the patients who need attention at the top of the physician's review queue.
4 fractures in one spine
Height loss against the height expected from neighbouring vertebrae. Real CT study, de-identified.
The problem
A patient has a CT for abdominal pain or a chest infection. The report answers that question. The same images show the spine, and in the spine there is a fracture. No one was looking for it, so no one mentions it.
Chest CT: one UK centre, patients over 55; fractures not correctly reported as osteoporotic fractures.1 Abdominal CT: one US centre, 2,041 consecutive adults scanned 2000–2007, before sagittal reconstructions were routine.2
It is a warning sign of osteoporosis, and it predicts the next break. Women with a spinal fracture face:
And yet
In the trials that established today's osteoporosis treatments, therapy cut new spinal fractures by roughly half or more. A patient whose fracture is never mentioned never gets that chance.
Every one of these fractures was visible on a scan that had already been taken, paid for and read.
Share of women with a new spinal fracture on X-ray in three randomised, placebo-controlled trials.789
The product
Six screens take a physician from a full archive to a reviewed, approved study.
Every CT is analysed automatically as it arrives. Patients with the most fractures rise to the top, so attention goes where it matters most.
The study opens in a multiplanar viewer: sagittal, coronal and axial planes, linked. Each vertebra is colour-graded for severity. Green is normal, orange moderate, red severe.
At T12 the front of the bone measures 12.6 mm. Judged against its neighbours it should be about 27.7 mm: a 54.5% loss of height. Every grade traces back to numbers a physician can check.
T8 is cut off by the edge of the scan. Instead of guessing, the system marks it purple, indeterminate, and states the reason, so a human reads it.
Any grade can be corrected in two clicks, then the study is approved. Each correction is stored beside the original AI grade, with the reviewer's name.
Workload, findings by severity and by vertebral level, and how often reviewers confirm the AI grade, on one screen.
Real screens from the portal, on real CT studies. Select an image to enlarge it.Swipe across an image to see the rest of it. Study identifiers were replaced before capture.
How it works
Spine Sight AI is designed to sit beside the imaging archive already in place. It speaks standard DICOM, reads without changing anything, and needs no internet connection.
Corrections are recorded with the reviewer's name and feed controlled retraining. A new model version is released only after it is checked on patients it has never seen.
A grade taken from a single slice depends on where that slice happens to fall. Spine Sight AI follows each vertebra through the whole scan, aligns to the curve of the spine, and measures its front, middle and back heights on parallel planes one millimetre apart.
The heights are compared with what the neighbouring vertebrae say this one should measure. The percentage lost sets the grade, using the thresholds radiologists already know from the Genant method.
A separately trained model then grades the same vertebra as a check. If the two disagree, the measurement is implausible, or the bone is cut off by the scan, the vertebra is marked indeterminate and goes to a reader.
Height loss: 12.6 mm measured at the front of T12 against 27.7 mm expected from its neighbours.
Calipers drawn on the image, heights in millimetres, the percentage lost and the rule that set the grade. Nothing rests on a score no one can inspect.
About one vertebra in seven is marked indeterminate, mostly at the edge of the scan, and each one states its reason. Those go to a human instead of being called normal.
Nothing is final until a physician approves it. Reviewer grades are stored separately from the AI output and are the reference for how the system is performing locally.
The whole pipeline has been run with every outbound network connection blocked. Images and results stay on the hospital's own server.
Evidence
Tested on patients the system had never seen, against grades given by radiologists. We publish the counts behind every percentage.
27 of 36 (75%; 95% CI 59–86%)
312 of 316 vertebrae without a fracture correctly left alone (95% CI 97–99.5%)
Positive predictive value 90.5% against the radiologists' grades
1,725 vertebrae in 160 CT scans; median boundary error against expert outlines 1 mm (median Dice 0.93)
The grading model was trained on public CT data and on scans from AUBMC patients. Six candidate models were compared, and the one in use was chosen on cross-validation before any test result was seen.
Fracture figures come from a separate public test set graded by radiologists; none of its patients were used in training. Counting mild fractures as well, the system found 38 of 64 (59%): most fractures called mild by eye measure under the 20% threshold. The numbering and outline figures use public benchmark scans, some of which the pretrained outlining model may have seen.
The system measures the shape of each vertebra from T1 to L5. Deciding the cause, such as a tumour, a Schmorl's node or a developmental shape, stays with the physician, who can record it in the review form.
Savings
Finding the first spinal fracture is what opens the door to preventing the next one. Here is what the next one costs, and what a hospital can expect to find.
direct medical cost per patient in the year after a hip fracture11
added medical cost in the year after any new osteoporotic fracture12
cost to US Medicare of the repeat fractures in patients who first fractured in 201612
that Medicare could have saved by preventing 5 to 20% of those repeat fractures12
Move the sliders. Every default comes from a published study or from our own test results, and you can change each one.
Treatment uptake is yours to set: in one UK service 69% of fracture patients were treated with a liaison service, against 19% in usual care.13 Cost of a repeat fracture: the added cost to US Medicare over 180 days.12
An estimate for each year's scans, built by chaining results from different studies and populations; your own figures will differ. It counts the cost of fractures avoided, before the cost of treatment and of the software, and the saving falls to whoever pays for fracture care.
Questions
No. It is a second reader for a finding that is easy to miss when the scan was ordered for something else. The physician reviews every flagged study, can change any grade, and approves the result. Nothing is final without that approval.
No. It works on routine chest, abdominal and pelvic CT, with or without contrast, on the thin axial series most protocols already produce. No extra scan, no added radiation and no change at the scanner.
No. Spine Sight AI is designed to run entirely on a server inside the hospital network. It needs no internet connection, loads nothing from outside, and the full pipeline has been tested with every outbound connection blocked.
Through standard DICOM. It queries the archive by accession number and retrieves the CT series, both read-only, so nothing in the archive is changed. Results open in a browser from a link that carries the accession number, which can be launched from the systems physicians already use.
One standard server. It does not need a GPU. Retrieval, outlining, measurement and grading take about two to three minutes per study on a workstation-class processor, and run in the background.
From measurement. Each vertebra's front, middle and back heights are measured on many parallel planes and compared with the height expected from its neighbours. Under 20% loss is normal, 20 to 25% mild, 25 to 40% moderate, and 40% or more severe. A separately trained model grades the same vertebra as a check.
It says so. A vertebra is marked indeterminate when it is cut off by the scan, when its outline fails quality checks, when the measurements are implausible, or when the two grading methods disagree. That is about 13 to 15% of vertebrae, and each one shows the reason.
No, by design. Corrections are stored as verified labels with the reviewer's name. Retraining is a separate, controlled step: the new version is checked on patients it has never seen, and it replaces the old one only if it performs better.
Contact us