Evidence
Where we are today.
Tested on patients the system had never seen, against grades given by radiologists. We publish the counts behind every percentage.
moderate and severe fractures found
27 of 36 (75%; 95% CI 59–86%)
specificity
312 of 316 vertebrae without a fracture correctly left alone (95% CI 97–99.5%)
flagged fractures confirmed
Positive predictive value 90.5% against the radiologists' grades
vertebrae correctly numbered
1,725 vertebrae in 160 CT scans; median boundary error against expert outlines 1 mm (median Dice 0.93)
How it was trained
The grading model was trained on public CT data and on scans from AUBMC patients. Six candidate models were compared, and the one in use was chosen on cross-validation before any test result was seen.
How it was tested
Fracture figures come from a separate public test set graded by radiologists; none of its patients were used in training. Counting mild fractures as well, the system found 38 of 64 (59%): most fractures called mild by eye measure under the 20% threshold. The numbering and outline figures use public benchmark scans, some of which the pretrained outlining model may have seen.